Mini Gerbil debugging – inconsistent connection or operation

Return to MG3 debugging

Symptom DetailLikely causes and their solutions (in recommended sequence)
Connection occurs, but inconsistent operationPoor connections into Mini Gerbil (one example: ribbon cable loose).  Check the following:
* Interference – other wires are near the USB cable (either inside the K40 or outside the K40). Keep other wires well away so the USB cable doesn’t pick up interference.
* Connection – use a different or shorter USB cable. Don’t use USB extension cables, couplers or hubs. If you absolutely have to, try cables/couplers specified for USB 3.0.
* Mounting – ensure MG3 is mounted via the provided plastic standoffs, rather than steel bolts on the K40.

Check for mechanical issues such as incorrect belt tension or obstructions

Virus scanner or other task interrupting your PC while laser is operating.  Disable and test.

PC is going into sleep mode during operation.  Change PC settings. Avoid running a laptop on battery power as this typically forces sleep mode.

PC USB system is going into sleep mode (note: this may not be visible to the user) Review PC settings, see https://apteryx.zendesk.com/hc/en-us/articles/360041367673-Disable-USB-Power-Management-Selective-Suspension

Faulty K40 grounding. See https://awesome.tech/k40-laser-grounding-instructions/

The stepsticks need to be adjusted for a lower or higher current, see https://awesome.tech/adjusting-stepsticks/